Agilent 6550 iFunnel Q-TOF LC/MS, G6550AA/G6550A

Usage

At PPB Analytical Incorporated, we offer researchers to have the chance to use the laboratory equipment in our lab at an affordable price. Book a QTOF capability with the Agilent 6550 iFunnel Q-TOF LC/MS, G6550AA/G6550A by calling us or sending an email.

Request A quote

To request a quote for the QTOF capability with the Agilent 6550 iFunnel Q-TOF LC/MS, G6550AA/G6550A, click here.

Product description

When you need maximum sensitivity, the 6550 iFunnel Q-TOF LC/MS system offers unprecedented detection levels in a high-resolution system. Ideally suited for proteomics research, where MS/MS is used consistently, the system delivers high precursor abundances, allowing you to acquire fragment data faster and reduce the duty cycle substantially.

The high sensitivity of the 6550 iFunnel Q-TOF LC/MS allows for highly repetitive quantitative analyses, allowing you to see compounds at trace level. The innovative combination of hexabore capillary and ion funnel-based entrance optics allows maximum capture of your analytes, making this the most sensitive Q-TOF LC/MS for your proteomics research.

Features
  • Achieve femtogram-level sensitivity with Agilent iFunnel technology, which increases ion transfer for the lowest detection levels of any high-resolution LC/MS instrument
  • Benefit from a high abundance of precursors that allow a fast duty cycle of fragmentation spectra, for more effective proteomics research
  • Avoid interferences and make high-resolution quantitation a reality with a powerful combination of accuracy and sensitivity
  • Gain confidence in your analysis with sub-ppm mass accuracy and resolution of 45k
  • See everything when screening or profiling with a wide in-spectrum dynamic range, complementing the system’s excellent detection limits and superb reproducibility
  • Agilent Infinity II UHPLC, 2D-LC, SFC, and GC-APCI chromatographic interfaces deliver exceptionally fast acquisition speeds
  • Features powerful MassHunter software for profiling, characterizing, identifying, and quantifying compounds in complex mixtures via high-definition MS and MS/MS
  • Use of the Agilent All Ions MS/MS technique facilitates untargeted screening and quantitation

Data Sheet

Parameter Measure Specification
Mass range   m/z 100 – 10,000 extended mass range m/z 50 – 1,700 or 50-3,200 for both high resolution and extended dynamic range modes Quadrupole up to 4,000 m/z
Mass resolving power ·         Measured at m/z 322 after automatic tuning procedure

 

·         Measured at m/z 2,722 after automatic tuning procedure

  • Greater than 25,000 at 322 m/z
  • Greater than 42,000 FWHM at 2,722 m/z
Sensitivity, MS mode, electrospray on-column, 400 µL/min fl ow rate 1 pg LC/MS injection of reserpine signal-to-noise for the reserpine (M+H)+ at m/z 609.2807 while maintaining a resolution of 40,000 at m/z 2,722 in 4 GHZ mode 1,000:1 RMS
Sensitivity, MS/MS mode, electrospray on-column, 400 µL/min fl ow rate 1 pg LC/MS injection of reserpine signal-to-noise for most intense product ions (174, 195,397,448 m/z) while maintaining a resolution of 40,000 at m/z 2,722 in 4 GHZ mode 3,000:1 RMS 4,500:1 RMS for maximum sensitivity mode
Mass accuracy, MS mode, electrospray on-column, 400 µL/min Measured at the (M+H)+ ion of reserpine (m/z 609.2807) using an internal mass reference Better than 1 ppm RMS as measured from 10 repeat injections
Mass accuracy, MS/MS mode, electrospray on-column, 400 µL/min Product ion 397 m/z for reserpine Better than 2 ppm RMS on m/z 397 as measured from 10 repeat injections
Mass accuracy temperature stability, MS mode Temperature: 15 to 35 °C (59 to 95 °F) at constant temperature Maintain 1 ppm mass accuracy (variations < 3 °C from calibration temperature)
Dynamic range Intra-scan dynamic range on co-eluting components Up to 5 decades
Spectral acquisition rate, MS mode m/z 50 to 1,700 in MS mode while maintaining a resolution of 40,000 at m/z 2,722 in 4 GHZ mode 50 spectra/second
Spectral acquisition rate, MS/MS mode m/z 50 to 1,700 in MS/MS mode while maintaining a resolution of 40,000 at m/z 2,722 in 4 GHZ mode 50 MS/MS spectra/second
Positive to negative switching Complete cycle switching from positive to negative and positive modes – allows for stabilization time 1.5 seconds